• Company
  • Products
  • Consulting
  • Bibliography

Technical parameters

Probes

 

ICR H-field probe

  • Vertical measuring coil
  • shielded measuring coil

 

 

Probe type

Inside diameter

Measurement range

Resolution

ICR HV100-27

100 µm

1.5 MHz - 6 GHz

60 µm

ICR HV100-6

2.5 MHz - 6 GHz

ICR HV150-27

150 µm

1.5 MHz - 6 GHz

80 µm

ICR HV150-6

2.5 MHz - 6 GHz

ICR HV250-75

250 µm

500 kHz - 2 GHz

110 µm

ICR HV250-6

2.5 MHz - 6 GHz

ICR HV500-75

500 µm

200 kHz - 1 GHz

300 µm

ICR HV500-6

2 MHz - 6 GHz

 

The particular of the probes ICR HVxxx 27/-75 are profitable in the lower frequency range.

Compare:  ICR product overview

 



 

ICR H-field probe

  • Horizontal measuring coil
  • shielded measuring coil

 

 

Probe type

Indide diameter

Measurement range

Resolution

ICR HH100-27

100 µm

1.5 MHz - 6 GHz

70 µm

ICR HH100-6

2.5 MHz - 6 GHz

ICR HH150-27

150 µm

1.5 MHz - 6 GHz

100 µm

ICR HH150-6

2.5 MHz - 6 GHz

ICR HH250-75

250 µm

500 kHz - 2 GHz

150 µm

ICR HH250-6

2.5 MHz - 6 GHz

ICR HH500-75

500 µm

200 kHz - 1 GHz

300 µm

ICR HH500-6

2 MHz - 6 GHz

 

The particular of the probes ICR HHxxx 27/-75 are profitable in the lower frequency range.

Compare:  ICR product overview

 



 

ICR E 150, E-field probe

  • Horizontal electrode
  • Dimensions of electrode: 150 x 35 µm
  • Measurement range: 7 MHz - 3 GHz
  • Resolution: 65 µm


 


Amplifier


Gain:

30 dB up to 3 GHz

20 dB up to 6 GHz

RF output:

50 Ohm

-1 dB Compression point:

120 dBµV

IP3:

27.0 dBm

Noise factor:

4.5 dB

Auxiliary power:

12 V +/-10%

 

 


BT 706 Bias-Tee


Power is supplied to the integrated amplifier via the bias-tee.This BT 706 has an impedance of 50 ohm and stabilises the current (9 V, 100 mA) for the ICR probe.

 

Frequency range:

500 KHz to 6 GHz

Connection:

SMA connector

Plug-in power supply unit

12 V / 70 mA


Note:

The bias tee from LANGER EMV-Technik features integrated voltage stabilisation at 9 V.

The ICR probes can thus only be used with this bias tee.




Properties and setup

ICR probe body

back