Quality verification for ICR Near-field microprobes

After the production each ICR-Microprobe has to be calibrated. Therefor the frequency response is measured above a defined stripeline and calibration data are created.

 

Measurement setup

Technical parameters:

 

Stripline

Width:

25 µm

    

Spectrum analyzer

TG-Level:

100 dBµV

Length:

2 cm

Ref.-Level:

100 dBµV

Closure:

50 Ohm

RBW:

1 MHz

Measure distance to probe:

20 µm

VBW:

100 KHz


Result:

For each delivered micro probe achieve our customers a certificate with the measured characteristic curves for the respective maximum and minimum. [Probe types] Both curves describes the quality of a micro probe. Furthermore  a correction file is created. With this file it is possible to measure with each delivered micro probe applying to a reference level the same. The required calculation could be done in ChipScan or Excel.

 

Probe characteristic (frequency response of the ICR-Near-field Microprobe)
Measuring result of the maximum above the same stripeline with the calibration data